文章摘要
张朝阳,刘海涛,许凯,杨锐,汤龙,孙会敏.聚乙二醇400分子量测定方法比较研究[J].中国药事,2018,32(8):1102-1107
聚乙二醇400分子量测定方法比较研究
Comparative Study on the Methods of Molecular Weight Determination of Polyethylene Glycol 400
投稿时间:2018-05-15  
DOI:10.16153/j.1002-7777.2018.08.015
中文关键词: 聚乙二醇400  高效凝胶色谱法  分子量  分子量分布
英文关键词: polyethylene glycol 400  high performance gel permeation chromatography (HPGPC)  molecular weight  distribution of molecular weight
基金项目:十三五国家新药创制-化药制剂质量关键技术评价研究(编号2015ZX09303001)
作者单位E-mail
张朝阳 中国食品药品检定研究院, 北京 100050  
刘海涛 河南省食品药品检验所, 郑州 450008  
许凯 中国食品药品检定研究院, 北京 100050  
杨锐 中国食品药品检定研究院, 北京 100050  
汤龙 中国食品药品检定研究院, 北京 100050  
孙会敏 中国食品药品检定研究院, 北京 100050 sunhm@126.com 
摘要点击次数: 2144
全文下载次数: 1207
中文摘要:
      目的:建立高效凝胶渗透色谱法(HPGPC)测定聚乙二醇400(PEG400)分子量及其分布,并与《中国药典》所用的端基分析滴定法比较。方法:采用Shodex OHpak SB-802.5HQ与Shodex OHpakSB-803HQ色谱柱串联分离,以0.1 mol·L-1硝酸钠(0.02%叠氮钠)为流动相,流速0.5 mL·min-1,柱温30℃,样品浓度2 mg·mL-1,采用示差折光检测器进行测定。结果:HPGPC测定的聚乙二醇400的分子量结果与《中国药典》2015年版所用的端基分析法测得的平均分子量结果接近,采用不同厂家色谱柱、对照品可影响PEG400的分子量测定结果,选用Shodex OHpak SB色谱柱串联测定聚乙二醇400重均分子量的范围为407~440,分子量分布宽度1.0~1.1,该方法的精密性RSD为0.2%,重复性为2.89%,24 h内样品稳定性RSD为0.66%。结论:经方法学验证,HPGPC法准确、重复性好,可用于聚乙二醇400分子量及其分布测定。
英文摘要:
      Objective:To establish the method of high performance gel permeation chromatography (HPGPC) to determine the molecular weight of polyethylene glycol 400 (PEG400) and its distribution and to compare it with the end-based titration method in the Chinese Pharmacopoeia. Methods:Shodex OHpak SB-802.5HQ and Shodex OHpak SB-803HQ columns were used to separate the samples. 0.1 mol·L-1 NaNO3 (0.02%NaN3) was used as the mobile phase, the flow speed was 0.5 mL·min-1, the column temperature was set at 30℃, and the sample concentration was 2 mg·mL-1, The refractive index detector was used to measure the results. Results:The results of the molecular weight of PEG400 measured by HPGPC was similar to those of the average molecular weight measured by the end-based titration method in the Chinese Pharmacopoeia (2015 edition). Different columns and reference standards could affect the molecular weight results of PEG400. The molecular weight range was from 407 to 440 and the distribution index was from 1.0 to 1.1 by using the method of Shodex OHpak column in series. The precision of the method was 2.2%, the reproducibility of the method was 2.89%,and the RSD of stability within 24 hours was 0.66%. Conclusion:Validated by methodology, HPGPC method is proved to be accurate, repeatable and can be used to determine the molecular weight and the distribution of PEG400.
查看全文   查看/发表评论  下载PDF阅读器
关闭